HOW TO CALCULATE COLDSPOT FOR AUTOCLAVE VALIDATION?
Answer / krishna
usually drain point is the cold spot of the autoclave. however, perform the validation by distributing the therocapules across the chamber and check the lowest temperature recorded during the sterilisation cycle. this is the lowest temperature of the autoclave