Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...


why we are taking 89ml concentrated HCL for preparing 1N HCL

Answers were Sorted based on User's Feedback



why we are taking 89ml concentrated HCL for preparing 1N HCL..

Answer / jyoti temkar

Actual strength of commercial HCl is 36%.
Molecular wt of HCl is 36.5
Density is 1.18 g/ml
So wt of HCl for 1 M is 36.5/(36/100) for 1000 ml
As volume = mass/ density,
Volume of HCl for 1M Strength in 1000 ml will be
36.5*100/(36*1.18)
=89 ml

Is This Answer Correct ?    19 Yes 4 No

why we are taking 89ml concentrated HCL for preparing 1N HCL..

Answer / abcd

since we r using fuming hcl ,so that the strength of the acid decrease while open it to atmosphere, so that we take 89 ml instead of 85 ml.

Is This Answer Correct ?    1 Yes 0 No

why we are taking 89ml concentrated HCL for preparing 1N HCL..

Answer / abcd

since we r using fuming HCl , so that the strength of the acid decreases if we kept open the acid to atmosphere ,so that we take 89 ml instead of 85 ml.

Is This Answer Correct ?    0 Yes 0 No

Post New Answer

More Analytical Chemistry Interview Questions

why digestion require in icpms?

0 Answers  


One Raw material failed assay test is there any chance for failing purity of that material ?

7 Answers   Mylan,


Plz explain the formula in uv calibration =Absorbance of the sample ( x ) x 10 x 1000/Weight of potassium dichromate taken in mg

1 Answers  


How to set specification of assay, fisdolution and related substances?

0 Answers  


Why we are performing wavelength accuracy at lower and higher wavelengths in HPLC calibration

3 Answers  


why sucrose used in polarimeter calibration?

2 Answers   Reddy Labs, Sugar Factory,


In acid base titration, after complet consumption of HCl by NaOH, addition of next drop of NaOH will give pink colour with phenolpthalein indicator. Why this pink colour appear? Why not blue, black, green or anything elese?

3 Answers  


what is the meaning of stress study?

2 Answers   Reddy Labs,


why does used dry caffien in HPLC calibration?

0 Answers   Dishman,


some product having water content method with pyridine and ethylene glycol mixture instead of methanol? how they are selecting and methanol is not suitable solvent how ?

0 Answers   Micro Labs, VMSRF,


Through assay by hplc(anhydrous basis) product got 100.8% .How much the potency of that product.Moisture content is 1.7%,Sulfated ash is 0.02%,Total impurities are 0.3%

2 Answers  


which one is better separation in high pressure liquide chromatogram orlowpressure chromato gram in hplc

0 Answers   Hetero,


Categories
  • Organic Chemistry Interview Questions Organic Chemistry (342)
  • Inorganic Chemistry Interview Questions Inorganic Chemistry (123)
  • Analytical Chemistry Interview Questions Analytical Chemistry (1387)
  • Physical Chemistry Interview Questions Physical Chemistry (62)
  • General Chemistry Interview Questions General Chemistry (478)
  • Chemistry AllOther Interview Questions Chemistry AllOther (190)