Why we more noise in PDA detector compared to UV detector?

Answer Posted / anil

There is no Integration function in the conventional UV-Vis spectrophotometer which
accumulates the signal. For example, the total required time will be 1000 sec. for 1000 data
points and it takes 1 sec. to measure one datum. In this case, all 1000 data have the same signal
to noise ratio (S/N).
But in a PDA instrument which has a 1000 photodiode array, 1000 data points can be
measured in 1 sec. and it would take 1/1000 sec. to achieve the same result obtainable in 1 sec.
in a conventional instrument. Therefore, when the same sample is measured for 1000 sec in a
PDA instrument, the signal is accumulated and is 1000 times greater than when measuring for 1
sec. The noise will be 10001/2
. This means that the S/N ratio is improved by 10001/2
.
This resulting benefit of fast data acquisition is termed Felgett’s S/N Advantage or
Multichannel Advantage

Is This Answer Correct ?    10 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

In performance Check of GC Why Hexadecane Peak is Considered 

2564


Difference between the quantitative analysis and qualitative analysis?

634


What is the difference between purge septum flow and column flow in gas chromatography?

1850


How to choose the which salt is suitable for mobilephase

2663


If vendor having more imp than monograph then how to proceed? and how to set spec?

351






WHAT IS THE PARAMETER FOR THE SELECTION OF MOBILE PHASE IN HPLC?

656


USP methodology, EP methodology, IP methodology, among three if possible to use one methodology for qualify working standard to use USP, EP, IP ? Please explain...

1566


what is the structure of DOWEX-50

4931


what is the importance of peak purity in HPLC and how we can calculate through manul(not software)?

3998


how decide the clining method and cleaning method validation require for this perticular products?

311


acceptance criteria for lod & loq by standard deviation of response and slope??

2103


WHY given much noise PDA dector then VWD Dector

1512


What is gelatinization?

1186


How can I develope method of dissolution by HPLC OR UV

1895


How require to develop GC method? how to select diluent, GAS, column selection and other chromatographic conditions?

283