What is calibration procedure of gredient HPLC pump

Answer Posted / prerna

Gradient Programme for port A and C :
00 : 100
00 : 100
25 : 75
25 : 75
50 : 50
50 : 50
75 : 25
75 : 25
100 : 00
100 : 00

Above programme also run for port B and D

An increasing stepwise piramid will appear. Integrate that
peak and also split each step of the peak.

Find out its height and area of height

By using the formula ( which I don't remeber ), mean flow
of solvent at each step.
Flow : Limit
25 (24-26)
50 (49-51)
75 (74-76)
100 (100)

If the observed value is within the acceptance limit then
that pump passes the test.

Is This Answer Correct ?    10 Yes 10 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

Did anybody have method for acetyl cysteine effervescnce tablet

1816


what is the different in Total ash, sulphated ash, acid insoluble ash, alcoholic or non alcoholic ash?

9448


How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample

2546


what type of question will ask in the interview of lab chemist.

2895


how we can identify the impurity is coming below loq at transfering site?

391






how to qualify the impurity?

328


What if impurity area in control sample coming more as compared to LOQ level of impurity ?

1152


Why only Copper standard is used to calibrate Atomic Absorption spectrophotometer?

2151


in which situation ion pair require to use?

281


What is aggregate and fragments in SEC?

1589


how require to fix the linearity and accuracy cincentration range?

668


What is the difference between Discriminating media and DPDM(Dissolution Profile with Different Media)

6926


How to choose the which salt is suitable for mobilephase

2661


Why six unit used for precision?

811


What is the formula for relative diffrence for standard solution in solution stability in validation?

314