Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

how peak can be sharp in hplc when it is come flat

Answer Posted / vivek k

Sample solvent stronger than the mobile phase Dissolve
sample in mobile phase or at least reduce the strength of
the sample solvent as much as possible
Sample mass overload Reduce the amount (mass) of sample
injected.
Silanol interaction with amines • Reduce mobile phase pH to
< 3.0
• Increase mobile phase ionic strength.
25mM to 50mM recommended
• Add a competing amine to the mobile phase.
10 mM TEA is usually sufficient.
• Select a stationary phase with a lower silanol activity.
Adsorption of acids on silica • Increase salt
concentration in the mobile phase
25 mM to 50 mM is usually sufficient
• Reduce the pH of the mobile phase to < 3.0
• Add a competing organic acid.
1% acetic acid or 0.1% TFA is usually sufficient
Column void Replace the HPLC column.
Attempts to fill-in the void are seldom worth the effort.

Is This Answer Correct ?    10 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

if peak get problamatic then what require to do?

782


Tell me something about Vitamin A test method by HPLC

2658


why we use a particular hplc column for a particular compound give reasons?please

2733


why does used dry caffien in HPLC calibration?

2018


Why we use potassium dichromate in uv calibration Exact reason behind it??

1240


In Dissolution Test why limit is define Q+5% what is the role of +5%.

3083


In the HPLC Calibration done wavelength accuracy done between 200nm-280nm .but not done remaining 300-400nm not done ?

902


How require to develop GC method? how to select diluent, GAS, column selection and other chromatographic conditions?

715


How much is the standard area for glc analysis

2262


In Assy by potentio metry titration method how select MET U ,DET U, SET U,

2139


Why only 1.2 million lux hours require in photostability study?

1007


acceptance criteria for lod & loq by standard deviation of response and slope??

2572


How would you decide dissolution medium for NCE compound of class I drug

2235


what is partion and column chromatography

2368


If we have 5 strength which is not dose proportinate and excipients also diffrent in each strength then how we can proceed for Force degradation? and excipient are same but not dose propotinate the how FD?

788