How can you calculate manually "Slope" in pH calibration?
Answer Posted / chmreddy
A good electrode has a mV of 0 ± 30 mV in pH 7.
The slope should be 90 to 105%.
Example:
Readings
pH 4 : 4.29
pH 7 : 7.43
Convert pH 7 reading to mV
7 – 7.43 = -0.43 pH * 59.16 mV/pH = -25.44 (this value is okay because it is within 0 ± 30mV)
Convert pH 4 reading to mV
7 – 4.29 = 2.71 pH * 59.16 mV/pH = 160.32
Calculate slope
160.32 – (-25.44) = 185.76 mV/ 177.48 mV * 100 = 105 %
(this value is borderline as 105% is the upper limit)
| Is This Answer Correct ? | 13 Yes | 11 No |
Post New Answer View All Answers
why glutent are detected in the rice cereal baby food product even manufacturer claimed that they are using rice and milk only?we have using ELISA to do the test,and rice supposed not containing any glutent,rite?We already repeat the test so many times and it still detected.just wondering where the glutent came from?
What is dose dumping? why require to do?
if rsd failed then what require to do?
What is rs test why we are performing rs test?
how pda detector works over uv?
can we use the same detector in HPLC as well GC and what could be the differences we can find in the final chromato graph in any aspects?
What is the difference between purge septum flow and column flow in gas chromatography?
If we do accuracy at same concentration at which linearity planned,what is the need to do linearity separately?
why sre you used Potassium hydrogen phthalate in standarisation of 1N NaOH and 0.1 N Perchloric Acid?
EXPLAIN THE CALCULATION PART OF AN KF TITRATOR CALIBRATION WITH SUITABLE EXAMPLE
how the compound separate in coulunm,explain
How we can identify process related and degradation impurity in single method with short period?
Hi sir if any product monograph not given known impurities (may be 5 impurities) specifications then how we require to proceed for inhouse formulation?
How doing qbd practically?
How require to perform linearity as per ANVISA? What are the acceptance criteria?