Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

How can you calculate manually "Slope" in pH calibration?

Answer Posted / chmreddy

A good electrode has a mV of 0 ± 30 mV in pH 7.
The slope should be 90 to 105%.
Example:
Readings
pH 4 : 4.29
pH 7 : 7.43
Convert pH 7 reading to mV
7 – 7.43 = -0.43 pH * 59.16 mV/pH = -25.44 (this value is okay because it is within 0 ± 30mV)
Convert pH 4 reading to mV
7 – 4.29 = 2.71 pH * 59.16 mV/pH = 160.32
Calculate slope
160.32 – (-25.44) = 185.76 mV/ 177.48 mV * 100 = 105 %
(this value is borderline as 105% is the upper limit)

Is This Answer Correct ?    13 Yes 11 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

why glutent are detected in the rice cereal baby food product even manufacturer claimed that they are using rice and milk only?we have using ELISA to do the test,and rice supposed not containing any glutent,rite?We already repeat the test so many times and it still detected.just wondering where the glutent came from?

2788


What is dose dumping? why require to do?

964


if rsd failed then what require to do?

849


What is rs test why we are performing rs test?

1871


how pda detector works over uv?

746


can we use the same detector in HPLC as well GC and what could be the differences we can find in the final chromato graph in any aspects?

2193


What is the difference between purge septum flow and column flow in gas chromatography?

2593


If we do accuracy at same concentration at which linearity planned,what is the need to do linearity separately?

2313


why sre you used Potassium hydrogen phthalate in standarisation of 1N NaOH and 0.1 N Perchloric Acid?

9810


EXPLAIN THE CALCULATION PART OF AN KF TITRATOR CALIBRATION WITH SUITABLE EXAMPLE

2652


how the compound separate in coulunm,explain

2541


How we can identify process related and degradation impurity in single method with short period?

832


Hi sir if any product monograph not given known impurities (may be 5 impurities) specifications then how we require to proceed for inhouse formulation?

711


How doing qbd practically?

984


How require to perform linearity as per ANVISA? What are the acceptance criteria?

1349