If i performing one analysis on C8column then the same analysis done on C18 column then what will happen about Rt;Resulution;Peak shape Explain ?

Answer Posted / t reddy

consider ur performing combination drug.... if both the drugs hydrophilic means.

in c8 column the rt will be high compare to c18 and resolution between two peaks also will increase in c8.

i think peak shape won't affect based on this factor.

Reason - c8 is less hydrophobic than c18... because of less carbon.... c18 is more hydrophobic so the interaction between hydrophilic drug and high hydrophobic column(c18) is less, drug will elute quickly and RT will be less than c8 column and resolution also.

NOTE - this is just a possibility, u can still get the good Rt with c8 also by changing the various factors like mobile phase composition, flow rate, temperature etc

Is This Answer Correct ?    2 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

What is aggregate and fragments in SEC?

1589


If suppose 10 methods of dissolution given in pharmacopoeial for single content product then which method out of 10? or all 10 require to follow?

530


how require to select dissolution media? what is discrimination?

315


Which parameters require to do in tech transfer?

280


in hplc chromatogram started from left to right and in uv spectrum started from left to right

471






why cone formation during dissolution?

326


How will u perform degradation study in detectors other thann PDA in HPLC Methodology

2021


What is intact assay method development.....?

2008


What is the accrptance criteria in RSD for RS method precision on basis of impurity percentages?

313


WHAT IS THE USE OF STARTER ?????????

1911


what is definition of validation? which components are followed give detail?

1831


what is diffrence between specificity and selecivity?

320


in DMF having extra impurities and in api COA also having extra imp than USP or BP product then how require to proceed?

348


how will you do the prep for unstable componds?

1443


Why only Copper standard is used to calibrate Atomic Absorption spectrophotometer?

2147