Why caffeine as a reference standard used for calibration
of HPLC?
Answer Posted / srinivasarao yemineni
1. Maxima at 205 nm
2. Minima at 245 nm
3. Maxima at 273 nm
Since it is a universal known standard.
It is stable and easily available in the market.
But here comes the question how about the calibration of
wavelength accuracy at Visible range. In this case we can
use Erbium Perchlorate which will scan the whole range(200
nm to 600 nm).
It is stable and little costly but availability in the
market when compared to caffeine is litte tough.
Thanks for Marking this Answer
| Is This Answer Correct ? | 6 Yes | 1 No |
Post New Answer View All Answers
What is the calibration of uv process and preparation
How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample
how to get accurate result for Residue on evaporation in purified water
In IR spectra, generally a molecule vibrates without apply emr or ir light?
from where i get mortar pastle for glass bottle crush? it's required for testing of it?
how you start RS method development when for new product?
How do we get end points and how many end points are possible for citric acid and di-acid not theorotically answer should be given practically.
How require to perform linearity as per ANVISA? What are the acceptance criteria?
Which parameters require to do in tech transfer?
can i use hplc detector to uplc and why?
which one is better separation in high pressure liquide chromatogram orlowpressure chromato gram in hplc
What is gelatinization?
Why use only methanol,used determinationof moisture content not use other solvents
in gas chromatography what is the difference between gas flow rate and average linear velocity ?
My question about gas chromatography sulfur chemiluminsecence detector. I test unknown sample gas by GC-SCD (calibrated ) and the result of *H2S is 279 PPM , *but when I test the same sample with the GC-TCD (calibrated ) the value of *H2S is 2500 PPM . I'd like to inform you that both GCs are calibrated and have very good operation conditions with stable parameters . the question is if the sample gas with higher H2S over detection limits of SCD detector (1000 ppm). why the result it 279 ppm Best regards