how peak can be sharp in hplc when it is come flat
Answer Posted / vivek k
Sample solvent stronger than the mobile phase Dissolve
sample in mobile phase or at least reduce the strength of
the sample solvent as much as possible
Sample mass overload Reduce the amount (mass) of sample
injected.
Silanol interaction with amines • Reduce mobile phase pH to
< 3.0
• Increase mobile phase ionic strength.
25mM to 50mM recommended
• Add a competing amine to the mobile phase.
10 mM TEA is usually sufficient.
• Select a stationary phase with a lower silanol activity.
Adsorption of acids on silica • Increase salt
concentration in the mobile phase
25 mM to 50 mM is usually sufficient
• Reduce the pH of the mobile phase to < 3.0
• Add a competing organic acid.
1% acetic acid or 0.1% TFA is usually sufficient
Column void Replace the HPLC column.
Attempts to fill-in the void are seldom worth the effort.
Is This Answer Correct ? | 10 Yes | 0 No |
Post New Answer View All Answers
What is the principle of HPLC, GC, LCMS, GCMS,LC QQQ, GC QQQ, LC Q TOF and GC Q TOF. What are the applications and Specificity?
is it necessary to do HPLC calibration at wavelength 315nm if we are doing analysis at this wavelangth
Why sometime potassium ortho phosphate mix with acetone use for sample preparation for hplc
what is partion and column chromatography
if you have given one product then how you determine the impurity in that?
we are performed SOR for a particular product the limit is -6 to -10.We face particular bathes it is not meeting the specification.we performed diffrent instrument diffrent analyst we are getting diffrent diffrent results.solvent is dimethyl sulfoxide what could be the reasion
calibrtion procedue of LC-MS
from where i get mortar pastle for glass bottle crush? it's required for testing of it?
how require to select dissolution media? what is discrimination?
how to develop the icp ms method? Application of icp ms?
before starting analytical method valodation what you checking? and how giving preference to start validation?
What is shaking level in GC?
how to calibrate hplc & gc
We use hicrome 10 cxs column in hplc analysis....pls tell me.. what is the meaning of 10 cxs?
How to compare XRD graphs against standard and carry polymorpism study of API's by powder XRD method?