Why we more noise in PDA detector compared to UV detector?
Answer Posted / anil
There is no Integration function in the conventional UV-Vis spectrophotometer which
accumulates the signal. For example, the total required time will be 1000 sec. for 1000 data
points and it takes 1 sec. to measure one datum. In this case, all 1000 data have the same signal
to noise ratio (S/N).
But in a PDA instrument which has a 1000 photodiode array, 1000 data points can be
measured in 1 sec. and it would take 1/1000 sec. to achieve the same result obtainable in 1 sec.
in a conventional instrument. Therefore, when the same sample is measured for 1000 sec in a
PDA instrument, the signal is accumulated and is 1000 times greater than when measuring for 1
sec. The noise will be 10001/2
. This means that the S/N ratio is improved by 10001/2
.
This resulting benefit of fast data acquisition is termed Felgett’s S/N Advantage or
Multichannel Advantage
| Is This Answer Correct ? | 10 Yes | 1 No |
Post New Answer View All Answers
In BP2013, Loperamid HCl monograph. Assay by titration with 0.1N sodium hydroxide using hydrocloric acid 0.01N and reading the volume added between the 2 points of inflexion. I have a question that if the diluent solvent is ethanol is certainly consumed a amount of volume of titrant, so this volume must be eliminated on the result calculation or not apart from first point which is subtracted above.
In which situation we require to analytical method validation of excipient?
How to establish relative response factor for hplc. Why it is required?
why holium oxide prepared in 10% HCLO4 SOLUTION
For standardization of volumetric solutions, acceptance limit for concentration is considered as 10% of nominal concentration and %RSD as NMT 0.2% (for appropriate no. of replicates). But , if %RSD has not met, what should be further action?
What is the calibration of uv process and preparation
How require to develop GC method? how to select diluent, GAS, column selection and other chromatographic conditions?
What should be the level/ Percentage of Arsenic in Arsenic free zinc.
give clarity of linearity and range in method validation
What is column in chromatography?
what is classification of elemental impurities? what is risk assement in elemental impurities?
How would you decide dissolution medium for NCE compound of class I drug
why glutent are detected in the rice cereal baby food product even manufacturer claimed that they are using rice and milk only?we have using ELISA to do the test,and rice supposed not containing any glutent,rite?We already repeat the test so many times and it still detected.just wondering where the glutent came from?
How to do regeneration of Metacarb Pb plus column?
is it nessesary to do solution stability for 7 days?