Why we more noise in PDA detector compared to UV detector?
Answer Posted / anil
There is no Integration function in the conventional UV-Vis spectrophotometer which
accumulates the signal. For example, the total required time will be 1000 sec. for 1000 data
points and it takes 1 sec. to measure one datum. In this case, all 1000 data have the same signal
to noise ratio (S/N).
But in a PDA instrument which has a 1000 photodiode array, 1000 data points can be
measured in 1 sec. and it would take 1/1000 sec. to achieve the same result obtainable in 1 sec.
in a conventional instrument. Therefore, when the same sample is measured for 1000 sec in a
PDA instrument, the signal is accumulated and is 1000 times greater than when measuring for 1
sec. The noise will be 10001/2
. This means that the S/N ratio is improved by 10001/2
.
This resulting benefit of fast data acquisition is termed Felgett’s S/N Advantage or
Multichannel Advantage
| Is This Answer Correct ? | 10 Yes | 1 No |
Post New Answer View All Answers
For standardization of volumetric solutions, acceptance limit for concentration is considered as 10% of nominal concentration and %RSD as NMT 0.2% (for appropriate no. of replicates). But , if %RSD has not met, what should be further action?
in OSD forms require to use gas chromatography?
what is %labelled amount in content uniformity of dosage unit and its calculation?
What is mean by PDR?
what is turbidimetric titration?Give 2 examples.
What is a difference between potency and purity?
what is the principle involved in turbidometry and nephelometry ?
if tech transfer fails whay require to do?
Why use only methanol,used determinationof moisture content not use other solvents
How do we fix the sample concentaryion in hplc method development. What is the basis?
in which situation require to use paddle and basket?
pharma industries oriented
My question about gas chromatography sulfur chemiluminsecence detector. I test unknown sample gas by GC-SCD (calibrated ) and the result of *H2S is 279 PPM , *but when I test the same sample with the GC-TCD (calibrated ) the value of *H2S is 2500 PPM . I'd like to inform you that both GCs are calibrated and have very good operation conditions with stable parameters . the question is if the sample gas with higher H2S over detection limits of SCD detector (1000 ppm). why the result it 279 ppm Best regards
1)What's the meaning of Absorption,give a example. 2)What's the meaning of Adsorption,give a example. 2)what is the difference between Absorption and Adsorption.
How to do regeneration of Metacarb Pb plus column?