Answer Posted / ajinkya p.
In HPLC there must calibrations for Pump, Detector,injector
Pump :- Should calibrated by checking flow accuracy test, Pressure holding test.
Detector :- Should calibrated by checking Linearity test, Detector wavelength accuracy test
Auto sampler :- Should calibrated by Injector Accuracy test,
Carryover test, Injector precision test, carry over test
| Is This Answer Correct ? | 48 Yes | 2 No |