Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

we calibrate HPLC once in 6 months while iam doing
calibration i came to know that calibration was failed.
then how can i justify that the samples analyzed on the
system during this 6 months was wright? how can i forward in
to the next step for those batches?

Answer Posted / k l reddy

Need not to analyze all previous batches but you should
analyze 6 previous batches(recent batches of before
calibration)on different system(which is qualified) and
compare the results with old results. Fallow the same
prociser until the resukts matched with old results.

Is This Answer Correct ?    2 Yes 2 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

is it nessesary to do solution stability for 7 days?

816


why does used dry caffien in HPLC calibration?

2076


How to calculation of residual solvent methanol (3000 ppm) standard solution, and how many quantity used of test sample

3324


Related substance method equivalency on control sample or spiked sample?

820


what is the importance of peak purity in HPLC and how we can calculate through manul(not software)?

4533


What is mean by PDR?

1025


process of Diclofenac sodium,IP.

2284


How to start the dissolution development for unknown tab?

1734


in api coa contains only process impurities and in product coa contains degradation impurities?

733


In Assy by potentio metry titration method how select MET U ,DET U, SET U,

2237


acceptance criteria for lod & loq by standard deviation of response and slope??

2639


How we performed the force degradation for drug substance, is any specific guideline is available for each parameter(Acidic, basic, oxidation,heat)? what conditions you mentained for above parameters.

2359


Why acetonitrile and water are used as extraction solvent when analysis melamine? I thought they are miscible and won't be able to separate...

1750


what is %labelled amount in content uniformity of dosage unit and its calculation?

5299


how to develop the icp ms method? Application of icp ms?

846