how calibrate the hplc detector
Answer Posted / m.sivalingeswara rao
Hplc detector calibration to be done to know the acuracy of
the emitted light wave length to the set value as well as
to check the sensitivity of the detecotr to respond with
the concentration of analyte.
To check accuracy generally caffeine solution is used
because it shows maxima and minima at particular wave
lengths(i.e maxima at 206nm,273nm and Minima at 241nm,in
this range we can check accuracy of lamp emitted light).
Sensitivity by permorming linearity test with different
concetration of caffeine solution.
Reproducibility by injecting the replicate injection of
Caffeine solution and measuring the %RSD of the responses
| Is This Answer Correct ? | 19 Yes | 0 No |
Post New Answer View All Answers
How we can identify process related and degradation impurity in single method with short period?
cefoperozone and sulbactam inj. hplc test method
In the HPLC Calibration done wavelength accuracy done between 200nm-280nm .but not done remaining 300-400nm not done ?
from where i get mortar pastle for glass bottle crush? it's required for testing of it?
How can the GC or HPLC method is selected to determine the impurity profile in drug product?
if you get peak in blank then what require to do?
why glutent are detected in the rice cereal baby food product even manufacturer claimed that they are using rice and milk only?we have using ELISA to do the test,and rice supposed not containing any glutent,rite?We already repeat the test so many times and it still detected.just wondering where the glutent came from?
Why only Copper standard is used to calibrate Atomic Absorption spectrophotometer?
CAN WE DETERMINE THE PEAK PURITY IN GAS CHROMATOGRAPHY ANALYSIS?
How to calculate coreletion coefficient
Why use only methanol,used determinationof moisture content not use other solvents
in OSD forms require to use gas chromatography?
How to determine water content of bis tetrazole amine mono ammonium salt
why we should take dst factor for below 1%moisture samples
How to do regeneration of Metacarb Pb plus column?