how calibrate the hplc detector

Answer Posted / m.sivalingeswara rao

Hplc detector calibration to be done to know the acuracy of
the emitted light wave length to the set value as well as
to check the sensitivity of the detecotr to respond with
the concentration of analyte.
To check accuracy generally caffeine solution is used
because it shows maxima and minima at particular wave
lengths(i.e maxima at 206nm,273nm and Minima at 241nm,in
this range we can check accuracy of lamp emitted light).
Sensitivity by permorming linearity test with different
concetration of caffeine solution.
Reproducibility by injecting the replicate injection of
Caffeine solution and measuring the %RSD of the responses

Is This Answer Correct ?    19 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

how to calibrate hplc & gc

4022


How would you decide dissolution medium for NCE compound of class I drug

1690


In IR spectra, generally a molecule vibrates without apply emr or ir light?

1428


WHY given much noise PDA dector then VWD Dector

1494


If we have 5 strength which is not dose proportinate and excipients also diffrent in each strength then how we can proceed for Force degradation? and excipient are same but not dose propotinate the how FD?

323






for which product require to do content uniformity? what is limit of cu?

304


USP methodology, EP methodology, IP methodology, among three if possible to use one methodology for qualify working standard to use USP, EP, IP ? Please explain...

1548


What is the difference between purge septum flow and column flow in gas chromatography?

1825


How to start the dissolution development for unknown tab?

1134


In GC calibration,what is the procedure to calibrate the TCD (thermal counductivity detector)suggest ?

2703


Which parameters require to do in tech transfer?

262


which batch require to use for analytical method validation?

324


in which situation ion pair require to use?

271


what is impurity profile. how to interpret this impurity profile to a drug product or drug substance.

2188


what is the acceptance criteria for enteric coated tablets in 0.1n hcl validation in each parameter?

332