same sample on two preapration gives differnt result during
hplc analysis
Answer Posted / reet
there may be any one of the reasons:
1. Air bubble in the injector needle (needle wash 2-3 times0
2. carry over might be there.
3. random variation in the flow rate
4. Integration problem (chk the processing method)
5. Sample itself may a problem !
i. solubility issues
ii.sample(generally proteins may be sticking to the
walls of the sample vial and also may be getting
irreversibly stuck on the surface of the column)
iii. temperature change.
| Is This Answer Correct ? | 5 Yes | 0 No |
Post New Answer View All Answers
in which situation require to change rs specification?
function of detecter in hplc ,gc and spectroscopy? function of carrier gas in gc?
why digestion require in icpms?
Why perchloric acid used for potentiometric titration
I have compare C2H2-air and C2H2-N2O flame AAS on determination calcium. Both use same range of std to plot calibration curve. (2-6ppm) When i measure the sample with phosphate, KCl and LaCl, C2H2-N2O flame give false positive result, around 0.5ppm. When i measure the sample with phosphste, KCl and EDTA. C2H2-N2O flame also give 0.5ppm false positive. But both above mentioned sample would not give false positive when measured by C2H2-air flame. What is the reason?
what is different when impergnated silica plates are used in separation of azo dyes using column chromatography?
What is the accrptance criteria in RSD for RS method precision on basis of impurity percentages?
How require to perform linearity as per ANVISA? What are the acceptance criteria?
What is intact assay method development.....?
what is the disso medium for tropsiun chloride
How would you decide dissolution medium for NCE compound of class I drug
effect of pore size, pore volume, partical size, column length, carbon load on retention time? what is carbon load? what is the use?
what is mean by dissolution hydrodynamics?
how to calculate elemental impurities?
which one is better separation in high pressure liquide chromatogram orlowpressure chromato gram in hplc