same sample on two preapration gives differnt result during
hplc analysis
Answer Posted / reet
there may be any one of the reasons:
1. Air bubble in the injector needle (needle wash 2-3 times0
2. carry over might be there.
3. random variation in the flow rate
4. Integration problem (chk the processing method)
5. Sample itself may a problem !
i. solubility issues
ii.sample(generally proteins may be sticking to the
walls of the sample vial and also may be getting
irreversibly stuck on the surface of the column)
iii. temperature change.
| Is This Answer Correct ? | 5 Yes | 0 No |
Post New Answer View All Answers
how you fix the limits of impurities?
in stress study if your api not soluble in hcl naoh h2o2 then what require to do?
what is difference between UV - VISIBLE MODEL NO like 1600,1601,1700 etc ? plz explain me
CAN WE DETERMINE THE PEAK PURITY IN GAS CHROMATOGRAPHY ANALYSIS?
analytical method validation require to with respect to release specification or shelf life specification?
mode of absorption in alimentary canal?
how the compound separate in coulunm,explain
Identify problem faced when mass spectroscopy is used with HPLC system ?
[3/30, 13:29] Manoj P Venkatpurwar: How hplc column selection according to structure? How mobile phase buffer selection on molecule structure?
some product having water content method with pyridine and ethylene glycol mixture instead of methanol? how they are selecting and methanol is not suitable solvent how ?
in dissolution if one bowl got 70 percent 2nd bowl got 80 percent and 3rd bowl got 90 percent then how proceed?
Why acetonitrile and water are used as extraction solvent when analysis melamine? I thought they are miscible and won't be able to separate...
how require to interprit the method precision data of hptlc?
what is the principle of UV Vis spectroscopy, AAS, ICP OES,ICPAES, ICP-MS and FTIR
please explain about aluminium hydroxide assay