Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

same sample on two preapration gives differnt result during
hplc analysis

Answer Posted / reet

there may be any one of the reasons:
1. Air bubble in the injector needle (needle wash 2-3 times0
2. carry over might be there.
3. random variation in the flow rate
4. Integration problem (chk the processing method)
5. Sample itself may a problem !
i. solubility issues
ii.sample(generally proteins may be sticking to the
walls of the sample vial and also may be getting
irreversibly stuck on the surface of the column)
iii. temperature change.

Is This Answer Correct ?    5 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

in which situation require to change rs specification?

749


function of detecter in hplc ,gc and spectroscopy? function of carrier gas in gc?

2672


why digestion require in icpms?

776


Why perchloric acid used for potentiometric titration

2684


I have compare C2H2-air and C2H2-N2O flame AAS on determination calcium. Both use same range of std to plot calibration curve. (2-6ppm) When i measure the sample with phosphate, KCl and LaCl, C2H2-N2O flame give false positive result, around 0.5ppm. When i measure the sample with phosphste, KCl and EDTA. C2H2-N2O flame also give 0.5ppm false positive. But both above mentioned sample would not give false positive when measured by C2H2-air flame. What is the reason?

3540


what is different when impergnated silica plates are used in separation of azo dyes using column chromatography?

2025


What is the accrptance criteria in RSD for RS method precision on basis of impurity percentages?

830


How require to perform linearity as per ANVISA? What are the acceptance criteria?

1342


What is intact assay method development.....?

2668


what is the disso medium for tropsiun chloride

2263


How would you decide dissolution medium for NCE compound of class I drug

2264


effect of pore size, pore volume, partical size, column length, carbon load on retention time? what is carbon load? what is the use?

931


what is mean by dissolution hydrodynamics?

827


how to calculate elemental impurities?

1034


which one is better separation in high pressure liquide chromatogram orlowpressure chromato gram in hplc

2165