Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

injection volume accruacy test i calibration of hplc

Answer Posted / ravi

In this calibration water factor (0.99602)to be used to
convert the weight in volume (micro liter)
Calculation should be as follows

Mean volume = (W1-W2)/6*0.99602 = µl

Limit = +-2% of the injection volume

Is This Answer Correct ?    5 Yes 6 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

Why sometime potassium ortho phosphate mix with acetone use for sample preparation for hplc

1500


what you do when compound having dwel wavelength?

805


if identification threshold crosses the limits then what next step?

915


What is third generation HPLC Columns?

2033


calibrtion procedue of LC-MS

2758


We use Potassium Dichromate solution for the calibration of UV-Visible Spectrophotometer in UV region. My question is to calibrate visible region which solution could be used in photometric accuracy

1963


what are risk assement in the analytical qbd?

805


What if impurity area in control sample coming more as compared to LOQ level of impurity ?

1650


In Assy by potentio metry titration method how select MET U ,DET U, SET U,

2240


What is stray light in uv what are units for stray light

1720


My question about gas chromatography sulfur chemiluminsecence detector. I test unknown sample gas by GC-SCD (calibrated ) and the result of *H2S is 279 PPM , *but when I test the same sample with the GC-TCD (calibrated ) the value of *H2S is 2500 PPM . I'd like to inform you that both GCs are calibrated and have very good operation conditions with stable parameters . the question is if the sample gas with higher H2S over detection limits of SCD detector (1000 ppm). why the result it 279 ppm Best regards

1651


What is split ratio in Gc? Splitless? how requirr to select?

943


some product having water content method with pyridine and ethylene glycol mixture instead of methanol? how they are selecting and methanol is not suitable solvent how ?

2704


If suppose 10 methods of dissolution given in pharmacopoeial for single content product then which method out of 10? or all 10 require to follow?

1040


What is intact assay method development.....?

2757