How to calculate LOD & LOQ in HPLC Validation Method
Answer Posted / sunil.madhukar rao kulkarni
We can calculate LOD and LOQ values using linearity
graph,insted of sigma value STEYX can be considered for
more acuurate vale of LOD and LOQ
where STEYX : Standard error of the pridected Y-value for
each X in a regression
Final calulation given as below :
LOD = STEYX x 3/ Slope
LOQ = STEYX x 10 /Slope
Is This Answer Correct ? | 44 Yes | 10 No |
Post New Answer View All Answers
What is the principle of Thermal conductivity detector and FID?
why are measure gas flow " ml " in Gas chromatography
what is mean by dissolution biowaiver study?
How do we fix the sample concentaryion in hplc method development?
if your product is soluble in 0.1n hcl and water then which you choose as media from these 2 media?
Why we select scan range from higer wavelength to lower wavelength in uv visible spectroscopy and ftir spectroscopy ? .
why we are using hexane in calibration of number of drop per mL
how require to fix the linearity and accuracy cincentration range?
How to know the estimated LOQ concentration in ppm
why we use a particular hplc column for a particular compound give reasons?please
My question about gas chromatography sulfur chemiluminsecence detector. I test unknown sample gas by GC-SCD (calibrated ) and the result of *H2S is 279 PPM , *but when I test the same sample with the GC-TCD (calibrated ) the value of *H2S is 2500 PPM . I'd like to inform you that both GCs are calibrated and have very good operation conditions with stable parameters . the question is if the sample gas with higher H2S over detection limits of SCD detector (1000 ppm). why the result it 279 ppm Best regards
How to set analyticl specification for combination products?
in hplc chromatogram started from left to right and in uv spectrum started from left to right
Why sodium hydroxide used for maintain pH of phosphate buffer
A ha tc is actual defination of retention time?