How can we calibrate THe Auto Titrator ?
Answer Posted / krishnakumar
response of electrode- using ph 4.0 7.0 and 9.0 read the
millivolt reading and make a graph with ph vs mv .graph
slope is between 90 percent to 110percent
burette . by selecting 1 ml 2ml 5 ml 10 ml and 20 ml
titrant is water. check the weight of water in each step
and 1 percent limit is allowed from the selected volume
temperature- compare with calibrated thermometer at minimum
4 intervals .1 percent error is allowed from the calibrater
thermometer.correlation coefficient of temperature graph
actual and set value should be greater than 0.9
Is This Answer Correct ? | 12 Yes | 6 No |
Post New Answer View All Answers
Explain the function of mitochondria?
Define electrochemical series? : Electrochemistry Corrosion
What is acid anhydride of h2co3?
Define oxidation potential and reduction potential? : Electrochemistry Corrosion
Explain the chemical formula for dichromate?
Define chemical corrosion. What are the types of chemical corrosion? : Electrochemistry Corrosion
What kinds of structural changes accompany bond-breaking and bond-forming in olefin polymerization? : Polymer Chemistry
Explain what is meant by "c2 symmetry" and "mirror plane symmetry". Can a molecule with "mirror plane symmetry" ever be chiral? : Polymer Chemistry
Consider the polymerization of 1,2-dichloroethylene, h(cl)c=c(cl)h. What structural variations are possible in poly(1,2-dichloroethylene)? : Polymer Chemistry
What is the role of a catalyst? : Polymer Chemistry
Define the full form of hcl?
Observe the following number to four significant figures and express the result in standard exponential notation: 0.006543210?
What do you mean by redox reaction? : Electrochemistry Corrosion
what is meaning of wet analysis.
Name the 4 physical states of matter?