How can we calibrate THe Auto Titrator ?
Answer Posted / krishnakumar
response of electrode- using ph 4.0 7.0 and 9.0 read the
millivolt reading and make a graph with ph vs mv .graph
slope is between 90 percent to 110percent
burette . by selecting 1 ml 2ml 5 ml 10 ml and 20 ml
titrant is water. check the weight of water in each step
and 1 percent limit is allowed from the selected volume
temperature- compare with calibrated thermometer at minimum
4 intervals .1 percent error is allowed from the calibrater
thermometer.correlation coefficient of temperature graph
actual and set value should be greater than 0.9
| Is This Answer Correct ? | 12 Yes | 6 No |
Post New Answer View All Answers
A polymer's structure influences its physical properties. Describe two structural variations that are possible for polypropylene but not for polyethylene. : Polymer Chemistry
What are some of the polymers that you encounter every day? Describe their physical properties. : Polymer Chemistry
Explain how do you extract ephedrine from a mineral block?
Define electro plating and electroless plating? : Electrochemistry Corrosion
Give an example of a polysaccharide?
Atactic polypropylene is always produced with the [cp2zr(r)]+ catalyst. Explain this on the basis of molecular-level interactions. : Polymer Chemistry
Define ph?
What is a tripod?
is is advisable to give a Pewter finish n Stainless Steel ? State the Advantages & disadvantages.
What are the methods used to control the corrosion? : Electrochemistry Corrosion
Name the element does the symbol au stand for?
Polymer chain growth can be terminated by b-hydride elimination or by reaction with h2. What is one advantage of the h2 reaction? : Polymer Chemistry
Explain the secondary structure of proteins?
What are the factors which affect corrosion? : Electrochemistry Corrosion
What is the nature of the bonding interaction between a metal and an olefin? : Polymer Chemistry