Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

injection volume accruacy test i calibration of hplc

Answer Posted / pratima bhandarkar

Connect union in to the HPLC.
Mobile phase: Methanol: Water (50:50)
Temperature : 25°C
Retention time :1min
Flow rate :0.5mL/min
Take one HPLC vial. Fill it with water. Weigh it (W1)
Make 6 injections with 50µl injection volume
Again weigh it (W2)
Mean volume = (W1-W2)/6*100 = µl

Limit = +-2% of the injection volume

Is This Answer Correct ?    8 Yes 3 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

what are the standard methods for HPLC?

2730


What should be the level/ Percentage of Arsenic in Arsenic free zinc.

2176


How to determine the EDTA content by potentiometry titration in Ceftriaxine sodium

2486


Why only 1.2 million lux hours require in photostability study?

959


what is diffrence between specificity and selecivity?

712


sop of a uv visible spectrophotometer double beam elico model

5859


what is difference between UV - VISIBLE MODEL NO like 1600,1601,1700 etc ? plz explain me

2626


If vendor having more imp than monograph then how to proceed? and how to set spec?

843


how to calculate elemental impurities?

984


i want model written test paper for the post of analyst in IOCL(indian oil corporation ltd) any one pls help me

2669


Why we use potassium dichromate in uv calibration Exact reason behind it??

1183


Tell me about analytical method validation in QC

6122


How to start the dissolution development for unknown tab?

1609


analytical method validation require to with respect to release specification or shelf life specification?

731


What is third generation HPLC Columns?

1927