How can v calibrate ftir ? Plz explain me

Answer Posted / a.sudhakar

FT-IR IS CALIBRATED BY USING POLYSTYRENE FILM

PUT THE POLYSTYRENE FILM INTO THE SAMPLE HOLDER
AND SCAN IT FIXED WAVE NUMBERS ARE OBTAINED THEY ARE
3060
2849
1942
1601
1583
1154
1028
ACCEPTANCE CRITERIA IS +/- 1 CM

RESOLUTION: THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 2870 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 2849.5 CM SHALL BE MORE THAN 0.33

THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 1589 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 1583 CM SHALL BE MORE THAN 0.08.

Is This Answer Correct ?    29 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

how require to interprit the method precision data of hptlc?

295


if identification threshold crosses the limits then what next step?

331


what is difference between UV - VISIBLE MODEL NO like 1600,1601,1700 etc ? plz explain me

2220


What is the formula for relative diffrence for standard solution in solution stability in validation?

303


HI,I CLEARED BOB CLERK EXAM. MY INTERVIEW WILL BE ON 9TH OCTOBER,2010.PLEASE SEND ME INTERVIEW QUESTIONS AND ANSWERS. THANK YOU.

1732






What is control room temperature and which guide line says?

1156


WHICH STANDARD WE USE IN SILICA ANALYSIS IN SILICA-GEL METHOD FOR SOLID SAMPLE

1617


WHAT IS THE PARAMETER FOR THE SELECTION OF MOBILE PHASE IN HPLC?

641


how many types of balance calibration test?

1721


What is rs test why we are performing rs test?

1301


how you fix the limits of impurities?

375


Why are use silicon oil mr?

740


How can we confirm the HPLC column is end-capped or not? Is it possible to identify by physical appearance?

2715


Many times I don't got a caffeine peak in calibration of hplc using guard column ❓

1161


Loss of linearity in tkn analysis, possible reasons?

798