Answer Posted / a.sudhakar
FT-IR IS CALIBRATED BY USING POLYSTYRENE FILM
PUT THE POLYSTYRENE FILM INTO THE SAMPLE HOLDER
AND SCAN IT FIXED WAVE NUMBERS ARE OBTAINED THEY ARE
3060
2849
1942
1601
1583
1154
1028
ACCEPTANCE CRITERIA IS +/- 1 CM
RESOLUTION: THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 2870 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 2849.5 CM SHALL BE MORE THAN 0.33
THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 1589 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 1583 CM SHALL BE MORE THAN 0.08.
Is This Answer Correct ? | 29 Yes | 1 No |
Post New Answer View All Answers
how require to interprit the method precision data of hptlc?
if identification threshold crosses the limits then what next step?
what is difference between UV - VISIBLE MODEL NO like 1600,1601,1700 etc ? plz explain me
What is the formula for relative diffrence for standard solution in solution stability in validation?
HI,I CLEARED BOB CLERK EXAM. MY INTERVIEW WILL BE ON 9TH OCTOBER,2010.PLEASE SEND ME INTERVIEW QUESTIONS AND ANSWERS. THANK YOU.
What is control room temperature and which guide line says?
WHICH STANDARD WE USE IN SILICA ANALYSIS IN SILICA-GEL METHOD FOR SOLID SAMPLE
WHAT IS THE PARAMETER FOR THE SELECTION OF MOBILE PHASE IN HPLC?
how many types of balance calibration test?
What is rs test why we are performing rs test?
how you fix the limits of impurities?
Why are use silicon oil mr?
How can we confirm the HPLC column is end-capped or not? Is it possible to identify by physical appearance?
Many times I don't got a caffeine peak in calibration of hplc using guard column ❓
Loss of linearity in tkn analysis, possible reasons?