FT-IR IS CALIBRATED BY USING POLYSTYRENE FILM
PUT THE POLYSTYRENE FILM INTO THE SAMPLE HOLDER
AND SCAN IT FIXED WAVE NUMBERS ARE OBTAINED THEY ARE
3060
2849
1942
1601
1583
1154
1028
ACCEPTANCE CRITERIA IS +/- 1 CM
RESOLUTION: THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 2870 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 2849.5 CM SHALL BE MORE THAN 0.33
THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 1589 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 1583 CM SHALL BE MORE THAN 0.08.
| Is This Answer Correct ? | 29 Yes | 1 No |
Post New Answer View All Answers
for which product require to do content uniformity? what is limit of cu?
CAN WE DETERMINE THE PEAK PURITY IN GAS CHROMATOGRAPHY ANALYSIS?
Why potassium dichromate is used in calibration of uv?
Identify problem faced when mass spectroscopy is used with HPLC system ?
What is intact assay method development.....?
what are the standard methods for HPLC?
some product having water content method with pyridine and ethylene glycol mixture instead of methanol? how they are selecting and methanol is not suitable solvent how ?
In the HPLC Calibration done wavelength accuracy done between 200nm-280nm .but not done remaining 300-400nm not done ?
if you get peak in blank then what require to do?
which one is better separation in high pressure liquide chromatogram orlowpressure chromato gram in hplc
how to select short coloum and long coloumn for new molecule
identification is for unknown? qualification for known? reporting for LOQ?
Which products will produce when acetic anhydride reacts with Ter-butanol, Sec-butanol, Iso butanol, Dimethylamine etc.
USP methodology, EP methodology, IP methodology, among three if possible to use one methodology for qualify working standard to use USP, EP, IP ? Please explain...
what is the acceptance criteria for enteric coated tablets in 0.1n hcl validation in each parameter?