Answer Posted / a.sudhakar
FT-IR IS CALIBRATED BY USING POLYSTYRENE FILM
PUT THE POLYSTYRENE FILM INTO THE SAMPLE HOLDER
AND SCAN IT FIXED WAVE NUMBERS ARE OBTAINED THEY ARE
3060
2849
1942
1601
1583
1154
1028
ACCEPTANCE CRITERIA IS +/- 1 CM
RESOLUTION: THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 2870 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 2849.5 CM SHALL BE MORE THAN 0.33
THE DIFFERENCE BETWEEN THE ABSORBANCE AT THE
ABSORBATION MINIMUM AT ABOUT 1589 CM AND THE ABSORBATION
MAXIMUM AT ABOUT 1583 CM SHALL BE MORE THAN 0.08.
| Is This Answer Correct ? | 29 Yes | 1 No |
Post New Answer View All Answers
what is the purge flow & how to calculate
What is the accrptance criteria in RSD for RS method precision on basis of impurity percentages?
If combination product how require to identify which imp is of which api?
effect of pore size, pore volume, partical size, column length, carbon load on retention time? what is carbon load? what is the use?
what is difference in ods I ods II is this is realy from sorbent / packing characteristics or from other means
all types of questions asked at the cipla interview
Colorimtry
How we can identify process related and degradation impurity in single method with short period?
from where i get mortar pastle for glass bottle crush? it's required for testing of it?
why we are using benzene, anyline in acetic anhdride assay titration method?
what parameters we will consider while developing a HPLC method and how we confirm our HPLC method is valid?
Why dissolution test is not performed in all of the products
Which parameters require to do in tech transfer?
Loss of linearity in tkn analysis, possible reasons?
what is the difference between potentiometric titration and karl fischer titration?