acceptance criteria for ph meter, hplc
detector,pump,injector,oven when we calibrate.
Answer Posted / saptale maruti
Flow rate ± 0.2% of set value
Flow rate consistancy RSD NMT1.0%
Compositional Accuaracy : Difference in Absorbance is 0.01AU
Difference in time NMT 20SEc
Delay Volume NMT 4.5 (For 4 ml Delay Volume System)
Injection Volume Accuracy is : Average volume for 18
injection is 19.6 to 20.4µl
Linearity 0.999
Wavelenght Accuracy Maxima 273± 2 nm
Minima 205 and 245± 2 nm
Oven Temperature 2 degree from set temp.
For Ph ± 0.02
Is This Answer Correct ? | 7 Yes | 2 No |
Post New Answer View All Answers
how can give the expiry period and restadardisation of volumetric solution
What is control room temperature and which guide line says?
what is the purge flow & how to calculate
In Dissolution Test why limit is define Q+5% what is the role of +5%.
how require to select dissolution media? what is discrimination?
how you start the solubility study?
Which parameters require to do in tech transfer?
in dissolution why pool sample needed? in which type of drug pool sample need?
Why we used in n-butyl acetate water content terminology while in ethyl acetate we used moisture content terminology?
which are the diffrent batches in the pharmaceuticals?
We use Potassium Dichromate solution for the calibration of UV-Visible Spectrophotometer in UV region. My question is to calibrate visible region which solution could be used in photometric accuracy
Which parameter require to do for analytical method equivalency?
why require to do water content for drug product?
If inhouse hplc related substance method is completly diffrent from Usp for finished proďuct with diffrent impurities then how require to prove method equivalecy?
i want model written test paper for the post of analyst in IOCL(indian oil corporation ltd) any one pls help me