Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

acceptance criteria for ph meter, hplc
detector,pump,injector,oven when we calibrate.

Answer Posted / saptale maruti

Flow rate ± 0.2% of set value
Flow rate consistancy RSD NMT1.0%
Compositional Accuaracy : Difference in Absorbance is 0.01AU
Difference in time NMT 20SEc
Delay Volume NMT 4.5 (For 4 ml Delay Volume System)
Injection Volume Accuracy is : Average volume for 18
injection is 19.6 to 20.4µl
Linearity 0.999
Wavelenght Accuracy Maxima 273± 2 nm
Minima 205 and 245± 2 nm
Oven Temperature 2 degree from set temp.

For Ph ± 0.02

Is This Answer Correct ?    7 Yes 2 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

What is the calibration of uv process and preparation

1942


please explain about aluminium hydroxide assay

1093


If combination product how require to identify which imp is of which api?

791


if you given one product then which tests you will perform?

665


could negative ions be produced by bombardment process in mass spectrometry?

1723


Why to use propyl paraben in HPLC gradient calibration

7320


For standardization of volumetric solutions, acceptance limit for concentration is considered as 10% of nominal concentration and %RSD as NMT 0.2% (for appropriate no. of replicates). But , if %RSD has not met, what should be further action?

1840


in OSD forms require to use gas chromatography?

778


What is similarity factor?

1148


on saturation solubility study data how we can find out the bcs class of drug?

829


what you do when compound having dwel wavelength?

704


Which type of column should i use to check the purity of high molecular weight protein using HPLC reverse phase column chromatography? Hi everyone. I wanna to check the purity of high molecular weight protein (collagen) with MW of ~130 kDa using a HPLC. I know C18

1858


how require to interprit the method precision data of hptlc?

761


how we can identify the impurity is coming below loq at transfering site?

810


Loss of linearity in tkn analysis, possible reasons?

1256