What are the limits for uniformity dosage content for tablets and capsules
Answer Posted / viswanadham.dama
Uniformity of dosage units are determined from two ways.
1.Content Uniformity
2.Weight variation
Content Uniformity is applied when tablet or capsules having
25mg or below and
when active pharmaceutical ingradient 25% weight present in
the total tablet weight.
weight variation test is versa.
Acceptance critiria for uniformity of Dosage units are for
content uniformity Acceptance value cannot be more than 15.
calculation of acceptance value is
A.V=(X-M)+K.S
where X=Mean value of 10 units
M=98.5 or 101.5
K=2.4(constant value) for 10 units
2.0(constant value) for 30 units
S=Standard Deviation value
in the above expression there are three cases
a)A.V=K.S(when X value lies 98.5 to 101.5)
b)A.V=(X-M)+K.S(when X value is more than 101.5)
c)A.V=(M-X)+K.S(when X value is less than 101.5)
| Is This Answer Correct ? | 34 Yes | 13 No |
Post New Answer View All Answers
why octanol used to determine the partition coefficient ?
What is the difference between Discriminating media and DPDM(Dissolution Profile with Different Media)
how require to select dissolution media? what is discrimination?
in each media we require to use SLS? how to proceed?
can we use the same detector in HPLC as well GC and what could be the differences we can find in the final chromato graph in any aspects?
if you have given a blank water how you analyse by general analytical techniques?
if tech transfer fails whay require to do?
we are performed SOR for a particular product the limit is -6 to -10.We face particular bathes it is not meeting the specification.we performed diffrent instrument diffrent analyst we are getting diffrent diffrent results.solvent is dimethyl sulfoxide what could be the reasion
why glutent are detected in the rice cereal baby food product even manufacturer claimed that they are using rice and milk only?we have using ELISA to do the test,and rice supposed not containing any glutent,rite?We already repeat the test so many times and it still detected.just wondering where the glutent came from?
From where require to take the RLD sample?
please explain about aluminium hydroxide assay
What is control room temperature and which guide line says?
How can I develope method of dissolution by HPLC OR UV
How we performed the force degradation for drug substance, is any specific guideline is available for each parameter(Acidic, basic, oxidation,heat)? what conditions you mentained for above parameters.
WHY given much noise PDA dector then VWD Dector