Answer Posted / nirav intas pharma
control of resolution in IR:
Take spectrum of polystyrine film,
the abs minima at 2870 cm- & abs maxima at 2849.5cm- is
greater than 0.33
the abs minima at 1589 cm- & abs maxima at 1583 cm- is
greater than 0.08
| Is This Answer Correct ? | 12 Yes | 0 No |
Post New Answer View All Answers
UV and PDA detector, which have less signal to noise ratio?
2. Two grams of Benzoic acid are dissolved in 200 ml of water and extracted with 200 ml of diethyl ether. The distribution coefficient of benzoic acid is 100, and its dissociation constant is 6.5 10-5. Calculate the distribution ratio (D) of benzoic acid at pH 2, 5, and 6. 3. Calculate D at pH 2 to 10 (1 unit apart) in the above problem, and plot D versus pH.