How to decide restandardisation due or retest or expiry of
Volumetric solutions prepared in wet chemistry lab?
Answer Posted / mr.bujji reddy
It is pre stability study of volumetric solutions. using a
standardisation.
For Eg: If you take 0.1 M Sulphuric acid, It is
standardised with sodium carbonate, we test it for 15 days
daily. the obtained molarity shuld be varied not more than
10 %. that is 0.09 to 0.11. then the expiry can be given
for 15 days.
Restandardisation period and retest period is same incase
of volumetric solutions.
Restandardisation date is decided based on the changes in
normality on testing everyday during pre stability studies.
The variation is little bit for 5 days. then we can set
restandardisation period for every 5 days.
| Is This Answer Correct ? | 32 Yes | 7 No |
Post New Answer View All Answers
how to selecet an exact coloumn for an new molecule development by hplc how to select exact salt as buffer for new molecule development by hplc what is the the process to select the mode of saparation of compoundes by hplc what is the use of ph of buffer what is use of buffer,ph,organic phase,ans methods how the molecules get saparated in coloumn,
how to get accurate result for Residue on evaporation in purified water
What is the Formula for coreletion coefficient in plhplc calibration
how to calculate elemental impurities?
Hi sir if any product monograph not given known impurities (may be 5 impurities) specifications then how we require to proceed for inhouse formulation?
How will u perform degradation study in detectors other thann PDA in HPLC Methodology
what is the acceptance criteria for enteric coated tablets in 0.1n hcl validation in each parameter?
what is the principle of UV Vis spectroscopy, AAS, ICP OES,ICPAES, ICP-MS and FTIR
Which parameter require to do for analytical method equivalency?
From where require to take the RLD sample?
what is the purge flow & how to calculate
Which products will produce when acetic anhydride reacts with Ter-butanol, Sec-butanol, Iso butanol, Dimethylamine etc.
Why to use propyl paraben in HPLC gradient calibration
Why we select scan range from higer wavelength to lower wavelength in uv visible spectroscopy and ftir spectroscopy ? .
if your product is soluble in 0.1n hcl and water then which you choose as media from these 2 media?