What is calibration procedure of gredient HPLC pump
Answer Posted / prerna
Gradient Programme for port A and C :
00 : 100
00 : 100
25 : 75
25 : 75
50 : 50
50 : 50
75 : 25
75 : 25
100 : 00
100 : 00
Above programme also run for port B and D
An increasing stepwise piramid will appear. Integrate that
peak and also split each step of the peak.
Find out its height and area of height
By using the formula ( which I don't remeber ), mean flow
of solvent at each step.
Flow : Limit
25 (24-26)
50 (49-51)
75 (74-76)
100 (100)
If the observed value is within the acceptance limit then
that pump passes the test.
| Is This Answer Correct ? | 10 Yes | 10 No |
Post New Answer View All Answers
1.What is the difference between method validation and method verification 2.Which guidelines proposed to method transfer
Explain the relations between number of carbon atoms in alkanes and retention time ?
cefoperozone and sulbactam inj. hplc test method
Why six unit used for precision?
What is the formula for relative diffrence for standard solution in solution stability in validation?
How to compare XRD graphs against standard and carry polymorpism study of API's by powder XRD method?
all types of questions asked at the cipla interview
WHAT IS THE PURPOSE WE USE HCL, CHOPPER, BURNER, MONOCHROMATOR in AAS? PLEASE EXPLAIN BRIEFLY
What is similarity factor?
if instrument calibration fails then what require to do?
in which situation require to take incident in validation?
how decide the clining method and cleaning method validation require for this perticular products?
what is mean by peak intigrity?
in hplc chromatogram started from left to right and in uv spectrum started from left to right
In IR spectra, generally a molecule vibrates without apply emr or ir light?