why we have to store the pH electrode in 4.01 buffer ?
Answer Posted / ashok maduri
Always keep your pH electrode moist. We recommend that you store your electrode in storage solution. If
storage solution is not available, use either a pH 4 or 7 buffer solution. DO NOT store electrode in
distilled or deionized water this will cause ions to leach out of the glass bulb and render your electrode
useless.
After storage, you may notice white KCl crystals forming outside your electrode. This will not interfere
with measurements. Simply rinse the electrode and blot dry before use.
| Is This Answer Correct ? | 3 Yes | 0 No |
Post New Answer View All Answers
is it necessary to do HPLC calibration at wavelength 315nm if we are doing analysis at this wavelangth
Please give idea about method development for cleaning method and how maco level establish
what is difference in ods I ods II is this is realy from sorbent / packing characteristics or from other means
If we have 5 strength which is not dose proportinate and excipients also diffrent in each strength then how we can proceed for Force degradation? and excipient are same but not dose propotinate the how FD?
What is intact assay method development.....?
What is diffrence between extractable volume and deliverable volume? Answer pls
How to know the estimated LOQ concentration in ppm
what is the importance of peak purity in HPLC and how we can calculate through manul(not software)?
What is Band broading effect explain
how to calibrate hplc & gc
effect of pore size, pore volume, partical size, column length, carbon load on retention time? what is carbon load? what is the use?
How to determine water content of bis tetrazole amine mono ammonium salt
what is the principle of UV Vis spectroscopy, AAS, ICP OES,ICPAES, ICP-MS and FTIR
why salisylic acid not using now days for disdolution calibration?
we are performed SOR for a particular product the limit is -6 to -10.We face particular bathes it is not meeting the specification.we performed diffrent instrument diffrent analyst we are getting diffrent diffrent results.solvent is dimethyl sulfoxide what could be the reasion