is there any material other than caffene to calibrate hplc
detector
Answer Posted / atom4chem
To check ultraviolet region (190-400nm) WE use Caffeine, Erbium perchlorate and Benzophenone. Caffeine and Erbium perchlorate for wavelenth check while Benzophenone for system precision, detection and injection linearity.
| Is This Answer Correct ? | 0 Yes | 0 No |