is there any material other than caffene to calibrate hplc
detector

Answer Posted / atom4chem

To check ultraviolet region (190-400nm) WE use Caffeine, Erbium perchlorate and Benzophenone. Caffeine and Erbium perchlorate for wavelenth check while Benzophenone for system precision, detection and injection linearity.

Is This Answer Correct ?    0 Yes 0 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

what is mean by extactable and leachable study?

559


what is the acceptance criteria for enteric coated tablets in 0.1n hcl validation in each parameter?

334


In HPLC (Waters) software Empower1,2 &3 what is the updation for each version

2464


While performing TOC sst analysis Zero shift disabled & sample analysis zero shift enabled why?

2342


how require to set assay concentration for standard and sample?

360






why glutent are detected in the rice cereal baby food product even manufacturer claimed that they are using rice and milk only?we have using ELISA to do the test,and rice supposed not containing any glutent,rite?We already repeat the test so many times and it still detected.just wondering where the glutent came from?

2336


what is mean by nitrosamine impurities?

536


Which is the highly polar and highly non polar column in HPLC?

1769


CAN WE DETERMINE THE PEAK PURITY IN GAS CHROMATOGRAPHY ANALYSIS?

698


why sre you used Potassium hydrogen phthalate in standarisation of 1N NaOH and 0.1 N Perchloric Acid?

8902


I have compare C2H2-air and C2H2-N2O flame AAS on determination calcium. Both use same range of std to plot calibration curve. (2-6ppm) When i measure the sample with phosphate, KCl and LaCl, C2H2-N2O flame give false positive result, around 0.5ppm. When i measure the sample with phosphste, KCl and EDTA. C2H2-N2O flame also give 0.5ppm false positive. But both above mentioned sample would not give false positive when measured by C2H2-air flame. What is the reason?

2970


which one is better separation in high pressure liquide chromatogram orlowpressure chromato gram in hplc

1723


how to select short coloum and long coloumn for new molecule

1813


what type of question will ask in the interview of lab chemist.

2882


Why irmoisture balance is used for corrugated box moisture testing

1920