what is difference of uv and pda

Answer Posted / rajib hazarika

1. By PDA detector, we can measure the area/height
of particular sample at wavelengths ranging from
190 to 800 nm by a single injection( 3D scan is possible(Response Vs time Vs wavelength). Whereas in uv detector scanning is not possible.)
2. for unknown sample with unknown lambda maximum with a single injection PDA we can determine the lambda maximum but by UV several injection is required with different wavelength.
(for method validation, process validation, Research and development PDA is prefered as compared to UV
3. If proper elution is established with the help of multi step gradient mobile phase and column then analysis of mixture of compounds(with different Lambda max) can be possible by a single injection in PDA.

Is This Answer Correct ?    8 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

mode of absorption in alimentary canal?

1913


Qc interview questions

1242


what is the disso medium for tropsiun chloride

1791


What is stray light in uv what are units for stray light

1006


What is delay volume?

835






How require to perform linearity as per ANVISA? What are the acceptance criteria?

701


WHY pKa value is more important than pH value of the mobile phase in HPLC?

1246


in sampling of raw material which side is pasted sampled labels

1238


why should we perform dissolution PVT calibration only by UV spectrometer, not by HPLC ?

1173


What is similarity factor?

710


analytical method validation require to with respect to release specification or shelf life specification?

269


what is the importance of peak purity in HPLC and how we can calculate through manul(not software)?

3977


identification is for unknown? qualification for known? reporting for LOQ?

269


We use hicrome 10 cxs column in hplc analysis....pls tell me.. what is the meaning of 10 cxs?

1144


what is biorelivent dissolution media?

294