Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

In the test for control of absorbance in UV calibration why
do we use only potassium dichromate and what is purpose of
taking a specified amount of 57.0-63.0mg?

Answer Posted / kishor solanke (alkem)

In UV specrophotometer calibration during control of
absorbance paramater the K2Cr2O7 is taken 60mg±3mg because
the resulting solution gives the absorbance between 0.2 to
0.9 units which is the high performance & optimum range for
UV spectrophotometer.

Is This Answer Correct ?    14 Yes 2 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

in DMF having extra impurities and in api COA also having extra imp than USP or BP product then how require to proceed?

840


Please tell me about the pH of Polycaboxylic ether is it in the 5-6 range ever or more than 6

2609


Which is the highly polar and highly non polar column in HPLC?

2309


How to determine the EDTA content by potentiometry titration in Ceftriaxine sodium

2549


What is a difference between potency and purity?

1222


For limit test of heavy metals in BP, Method C require that the substance is ignited at a temperature not exceeding 800 °C. Why confines such the temperature?

4076


In HPLC (Waters) software Empower1,2 &3 what is the updation for each version

3032


how can give the expiry period and restadardisation of volumetric solution

2598


Home heating oil or furnace oil what is made from? Can (base oil plus diesel/kerosen) = home heating oil?

2187


How do we fix the sample concentaryion in hplc method development?

1144


if rsd failed then what require to do?

841


Which parameters require to do in tech transfer?

760


Why do we use KMnO4 in the test of control of obsorbance ? and why do we take specific quantity i.e 57-63mg?

2833


what is the origin to prepare standard operating procedure

2334


IS THERE ANY EQUIPMENT TO CHECK AND CALCULATE THE POLARITY OF A LIQUID?

2465