In the test for control of absorbance in UV calibration why
do we use only potassium dichromate and what is purpose of
taking a specified amount of 57.0-63.0mg?
Answer Posted / kishor solanke (alkem)
In UV specrophotometer calibration during control of
absorbance paramater the K2Cr2O7 is taken 60mg±3mg because
the resulting solution gives the absorbance between 0.2 to
0.9 units which is the high performance & optimum range for
UV spectrophotometer.
| Is This Answer Correct ? | 14 Yes | 2 No |
Post New Answer View All Answers
What is column in chromatography?
why we use a particular hplc column for a particular compound give reasons?please
How can I develope method of dissolution by HPLC OR UV
How we choose the mobile phase for method development.?
what is impurity profile. how to interpret this impurity profile to a drug product or drug substance.
What is the main difference between ODS1 and ODS2 Hplc column.
For limit test of heavy metals in BP, Method C require that the substance is ignited at a temperature not exceeding 800 °C. Why confines such the temperature?
4. Describe the operation of the Craig apparatus. Chapter.3 Equilibrium Processes in Separations
Why to use propyl paraben in HPLC gradient calibration
Home heating oil or furnace oil what is made from? Can (base oil plus diesel/kerosen) = home heating oil?
acceptance criteria for lod & loq by standard deviation of response and slope??
We use hicrome 10 cxs column in hplc analysis....pls tell me.. what is the meaning of 10 cxs?
from where i get mortar pastle for glass bottle crush? it's required for testing of it?
What is importance of pH in the Heavy metal test and How it effect on the test results?
CAN WE DETERMINE THE PEAK PURITY IN GAS CHROMATOGRAPHY ANALYSIS?