why do we use around only 60 mg of K2cr2o7 in the UV
calibration -control of absorbance test?
Answer Posted / kishor solanke
Because this resulting solution gives absorance in the
range between 0.2 to 0.9 absobance which is the optimum
range of absorbance for UV-spectrophotometer.
| Is This Answer Correct ? | 21 Yes | 1 No |
Post New Answer View All Answers