Golgappa.net | Golgappa.org | BagIndia.net | BodyIndia.Com | CabIndia.net | CarsBikes.net | CarsBikes.org | CashIndia.net | ConsumerIndia.net | CookingIndia.net | DataIndia.net | DealIndia.net | EmailIndia.net | FirstTablet.com | FirstTourist.com | ForsaleIndia.net | IndiaBody.Com | IndiaCab.net | IndiaCash.net | IndiaModel.net | KidForum.net | OfficeIndia.net | PaysIndia.com | RestaurantIndia.net | RestaurantsIndia.net | SaleForum.net | SellForum.net | SoldIndia.com | StarIndia.net | TomatoCab.com | TomatoCabs.com | TownIndia.com
Interested to Buy Any Domain ? << Click Here >> for more details...

Column selection in HPLC method development?

Answer Posted / satheesh kumar

Columns selection is main role for the separation of
compounds in method development.

1)Compounds solubility to be check
2)pKa of the compound

we know if the compound is acidic,fix pH nearer to pKa(+/-2)
Starts with ODS Column check the Compound RT,peak shape and
tailing criteria.
Generally HPLC Columns work between pH 2-8.If too acidic or
basic then columns packing will dissociate,
Some special columns are there to work in higher pH 8 to
namely X-Terra RP 18,Phenomenx Gemini C18.

Ensure what pH to be works for the concerned compound
select the columns accordingly

Is This Answer Correct ?    11 Yes 1 No



Post New Answer       View All Answers


Please Help Members By Posting Answers For Below Questions

what is mean by ambient temperature?

972


In performance Check of GC Why Hexadecane Peak is Considered 

3102


in which situation ion pair require to use?

876


what is lod and loq ?,why use k2cr2o7 , kcl h2so4 in uv calibration ?,why use benzophenone & caffene acetone in hplc calibration ?,what is leading peak in hplc ?why we do the calibration of limit of stry light in hplc & uv ?

3721


Ratio of Water and Acetonitrile(diluent) is not recommended to use in cleaning method validation,why?

3016


what is the origin to prepare standard operating procedure

2401


if you have given a blank water how you analyse by general analytical techniques?

2081


Ph meter can show more than 14 ph reading? Why ph range in between 1 to 14 only?

793


what is the purge flow & how to calculate

2724


what is classification of elemental impurities? what is risk assement in elemental impurities?

804


if peak get problamatic then what require to do?

856


How to do regeneration of Metacarb Pb plus column?

2610


What is the acceptance criteria for moisture balance when calibrated with sodium sulfate

2087


for heavy metal test lead used in which form pure form or any other form for preparation of lead standard sol

6141


please explain about aluminium hydroxide assay

1176