In Analytical method development related substance method
when did follow w/w method when did follow area
normalisation method
Answer Posted / m krishnamurthy
If all impurities havivg RRF value in the range of 0.8 to
1.2, Then we will follw area normalisation method. If RRF
is more than 3, then weshould follow w/w method .
| Is This Answer Correct ? | 23 Yes | 0 No |
Post New Answer View All Answers
Why we select scan range from higer wavelength to lower wavelength in uv visible spectroscopy and ftir spectroscopy ? .
For limit test of heavy metals in BP, Method C require that the substance is ignited at a temperature not exceeding 800 °C. Why confines such the temperature?
why we use glass fiber filters use in some situation?
If suppose 10 methods of dissolution given in pharmacopoeial for single content product then which method out of 10? or all 10 require to follow?
Difference between hlaf and rlaf
What is split ratio in Gc? Splitless? how requirr to select?
how require to set assay concentration for standard and sample?
Why do we use KMnO4 in the test of control of obsorbance ? and why do we take specific quantity i.e 57-63mg?
Which type of column should i use to check the purity of high molecular weight protein using HPLC reverse phase column chromatography? Hi everyone. I wanna to check the purity of high molecular weight protein (collagen) with MW of ~130 kDa using a HPLC. I know C18
before starting analytical method valodation what you checking? and how giving preference to start validation?
sop of a uv visible spectrophotometer double beam elico model
How we performed the force degradation for drug substance, is any specific guideline is available for each parameter(Acidic, basic, oxidation,heat)? what conditions you mentained for above parameters.
In rs method development when we are going area normalization method to dilute standard method?
We use Potassium Dichromate solution for the calibration of UV-Visible Spectrophotometer in UV region. My question is to calibrate visible region which solution could be used in photometric accuracy
how to decide for one product require water content or LOD?