Why we use caffeine for HPLC calibration in Detector wave
ength accuracy test
Answer Posted / karunakar reddy > elico ltd
> 4 caffeiene having different wavelents it having 2
maximas
and 1 minima.
1st maxima is 205nm
2nd maxima is 273nm
minima is 245nm
and it is primary reference standard and also suggested in
pharmacopiea.