How to calibrate vertex flow meter?
How to calibrate mass flow meter?
Answer Posted / jigaranadkat
Vortex:
Practically it is a volumeteric flow measuring device and it is not possible to calibrate online in field in most of the cases (Provide you have a proper bypass arrangement and proovers in place) , hence it is to be sent to calibration LAB where it will be compared with MASTER meter.
This MASTER meter must have superior accuracy than UUC (Unit Under Calibration).
Usually the large volume flow is tough to establish at calibration LAB hence Gravymeteric method is used in which water is passed through UUC and compared with high accuracy Load-cell. Since water density is known and constant, volume flow is calculated and new K-factor of meter is calculated.
Mass-flowmeter / Coriolis meter : Same Above mentioned Gravymeteric method is used.
Hope this will help ..!!!
Jigar Anadkat
Is This Answer Correct ? | 0 Yes | 0 No |
Post New Answer View All Answers
hello sir/madam, I,Subramanian need 5yrs placement papers of officer trainee of Instrumentation technology.pls send to this mail id mani_nxtgen@yahoo.co.in please do the needful. with regards, subramanian.v
how to connect transmitter measuring sulfur process? 1-by using long impulse line or 2- by using coated short impulse line
Describe the function of Accumulator into the Pump
why we use impulse line of flow transmitter at 45 deg. angle both side of orifice plate.
i need some objective type questions in instrumentation
Explain why use mmwc unit?
Which is your favourite subject & explain any one topic of it.
Why humidity limit is 30 to 55 % for raw material storage?
WHAT IS TMXT IN TEMPERATURE FOUNDATION FIELDBUS.
What is feed factor and its Importance in cement industry? How we can solve feed factor?.
How will you calibrate PID controller? explain with example.
How addressable fire systems work? Pl. send details with architect.(Loop), also send details regarding Gas system. Pl. help me.Thanks
in natural gas indirect water bath heater,how to prevent the problem of backfire from firetube during ignition...becoz of this backfire performance of wbh is going low.
Can anybody tell me which type of question are asked in the interview for the FORMULATION(Pharma) as a instrumentation engineer, & question in building managment system.
Hook up resistance decade box to temp.transmitter.