Answer Posted / carl_ellis
Disconnect the field sensor.
Connect a calibrated/traceable temperature signal calibrator.
Apply LRV value.
Trim the 4.0mA output.
Apply the URV value.
Trim the 20.0mA output.
Check as many interim points as the calibration protocol requires and note the output and the deviation from the expected value.
Determine whether the transmitter is within spec from the deviation values seen.
If transmitter passes, reconnect the field sensor to the input.
Fill in the paperwork.
Carl Ellis
Measure First
| Is This Answer Correct ? | 9 Yes | 3 No |
Post New Answer View All Answers
what is the calibration procedure for magnetic flow meter @ onsite
selection of the zener barrier for the hazardeous area zone 0
Why mmwc unit is used?
how many types of viberation transmitter and how to calibrate and where use of vibration transmitter
How we calculate orifice plat size???
what is Best Among SINK Or SOURCE Type Design In Digital Signal, Either Iput Or Output? WHY?Can any one..
What is feed factor and its Importance in cement industry? How we can solve feed factor?.
What are the charcterstics of a P&ID controller block for Governor speed raise or lower for a load sharing functionality between 4 Gas turbines in droop mode.
what do u mean by compensation n pid tunning?in pid tunning what you tunned?is it possible 2 tune p i d
what are the diff. in servo and regulatory control?
all types of liquid analyzer principle with drawing?
How to check vortex flow meter is in good condition while we do not having calibration setup?
In Safe Guarding test what means by "Failed to Danger" ?
What are primary elements of measuring pressure?
Explain the working of an enraf level gauge.