why some samples are taking 500mg for water content and some
samples are taken 2g for water content?
Answer Posted / kishor solanke
Test Preparation— Unless otherwise specified in the individual monograph, use an accurately weighed or measured amount of the specimen under test estimated to contain 2-250 mg of water The amount of water depends on the water equivalency factor of the Reagent and on the method of endpoint determination. In most cases, the minimum amount of specimen, in mg, can be estimated using the formula:
FCV / KF
in which F is the water equivalency factor of the Reagent, in mg per mL; C is the used volume, in percent, of the capacity of the buret; V is the buret volume, in mL; and KF is the limit or reasonable expected water content in the sample, in percent. C is between 30% and 100% for manual titration, and between 10% and 100% for the instrumental method endpoint determination.
Is This Answer Correct ? | 8 Yes | 0 No |
Post New Answer View All Answers
What steps would you take to ensure responsible prescribing and prevent overuse or misuse of Alprazolam in your patients?
whem will be the exam or interview for the post of Drug inspector is conducted in jammu and kashmir??? the form was filled on starting of 2013.
How to select temperature of ODT in disintegration test
what is the calculation for getting the tablet weight and the punch size, Eg..if there is a 8/32 punch then how to measure tablet weight????
why some samples are taking 500mg for water content and some samples are taken 2g for water content?
1.why did you choose this career 2.what is pharmacists ideal job
why we use matrixing method in stability study and on what basis time points are selected?
pharmacovigilance opportunities in Hyderabad
How we define validity of any Reagent , Indicator and Volumetric Solutions
What are the advantages and disadvantages of working for a small & big pharmaceutical company?
how you can fix the known and unknown impurity limit for any drug substance
when we calibrate uv for control of wavelength we scan for 200 to 800 nm but we get the result only for some specific wavelength how we confirm for all other wavelength?
What do you mean by Bracketing and Matrixing in stability?
what is difference between usfda and mhra guidline
what is purpose of slow,medium,fast sampling in process validation ?