Answer Posted / ashish namdev, bina refinery
interface LT use in that application where two liquid mixed
in a tank. For calibration interface type LT these
instructions to be follows -
1. set upper & lower range.
2. set upper & lower density.
3. after then put weight according calibration data sheet
& note readings at every sample weights.
4. If the output have some error then give lower trim & upper
trim command to LT with weight at 0% & 100%.
5. recheck all value at different weights.
this LT work upon bouncy principal. which liquid will have
high density that will be make reference to LT.
Is This Answer Correct ? | 3 Yes | 3 No |
Post New Answer View All Answers
HOW TO CHOOSE FREAQUENCY IN FMCW TYPE RADAR INSTRUMENTS. DESIGN BASIS OR DESIGN CALCULATONS ?
What is intrinsic safety mean.
How to calculate mmwc values into tonnes?
Specific gravity.
HI, can anybody give details about GAS TREATMENT PLANT CONSTRUCTION,as INSTRUMENTATION ENGINEER what are the responsibilites.what are the instruments mainly instrumentation engineer will deal with
Good evening all I need a small favour regarding reliance paper i wanna probable sample papers(e&i) of reliance. if any of u have any idea about dis then plz mail me back.i belongs to electronics and instrumentation engg.so i need e&i papers. thanking u....
Define cryogenic?
please send the sample qustion paper for iocl written test ..
can we replace a slot sensor (inductive type) with limit switch (potenial free contact / no-nc type).
How to calibrate Toxic gas detector?
Manipulation of dead wt. tester double piston type.
Derive capacitance to 4-20ma conversion formula / relation in capacitance type pressure sensor?
please tell me what is a light up procedure of Cfbc boiler and what it's axuillaries of cfbc boiler?
what happen if i increase height of displacer type LT.accuracy or sensitivity are affected.why?
give some question relative instrument fitter?